Identification of an Impurity in Pravastatin by Apliccation of Collision-Activated Decomposition Mass Spectra
Andrej Kocijan,a Rok Grahek,a Lucija Zupančič-Kralj b
aLek
Pharmaceuticals d.d., Development Center Slovenia, Verovškova 57, SI-1526
Ljubljana, Slovenia
b Faculty of
Chemistry and Chemical Technology, Aškerčeva 5, SI-1000 Ljubljana, Slovenia
Absrtract
Pravastatin is produced by a two step fermentation. Mevastatin
and 6-epi pravastatin are the main impurities in the fermentation broth
as well as in final product. An unknown impurity with m/z
437 for (M-H)-
was detected in analysis of pravastatin sodium sample by reversed phase HPLC-MS.
CAD spectrum of the impurity was obtained using HPLC-MS/MS equipped with
negative APCI. The CAD mass spectra of impurity was similar with CAD mass
spectra of statins. The structure of an impurity was finally determined on the
basis of proposed fragmentation mechanisms for statins.
Keywords: HMG-CoA reductase inhibitors, statins, tandem mass spectrometry, impurity identification