Gregor Arh,1,* Simona Fras,1 Tomaž Polak,2 Božidar Žlender,2 Marjan Veber1 and Matevž Pompe1
1 Faculty of Chemistry and Chemical Technology, University of Ljubljana,
Aškerčeva 5, 1000 Ljubljana, Slovenia
2 Biotechnical Faculty, University of Ljubljana, Jamnikarjeva 101, 1000
Ljubljana, Slovenia
* Corresponding author: E-mail: gregor.arh@fkkt.uni-lj.si;
Fax: ++386 1 2419 220
Abstract
In the paper the modification of standard method SIST EN 1528 1-4: 1998: Method
D for measurements of organochlorine
pesticides (OCPs) in food products is presented. The modifications were made in
the extraction step in which cold
extraction technique was replaced by Soxhlet extraction. For the clean-up step
smaller columns were introduced and for
the concentration step a rotary evaporation was replaced by Kuderna-Danish
concentration technique. Introduced modifications
improved the efficiency of the procedure for samples with high fat content.
Recovery values for all analyzed
pesticides were over 60% and the reproducibility expressed as relative standard
deviation was in the range of 10%. The
method is suitable for the determination of OCPs in meat products with high
content of fat from low ppb concentration
range onward. The limits of detection for examined OCPs were in the range from
0.1 ppb to 2 ppb for lindane and α-endosulfan,
respectively.
Keywords: Organochlorine pesticides; soxhlet extraction; florisil; ECD; meat